Home / Recalls / Medical Devices
Smith & Nephew Inc. recalls Smith & Nephew RENASYS-F XL with Soft Port REF #: 66800797 (The RENASYS Soft Port Dressing Kits are used in conjuncti
Recalled March 19, 2014 · FDA recall Z-1159-2014 · Smith & Nephew
Hazard
Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo · FDA Class II
Units
15,754 indivdual units
Sold at
Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
Description
Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
Product
Smith & Nephew RENASYS-F XL with Soft Port REF #: 66800797 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
Manufacturer(s)
Smith & Nephew Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →