Home / Recalls / Medical Devices
Smith & Nephew, Inc. recalls TC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 6 RIGHT, STERILE R, REF 12 000 0011, S&N 7500266, Smith & Ne
Recalled April 27, 2016 · FDA recall Z-1541-2016 · Smith & Nephew
Hazard
The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery. · FDA Class II
Units
4 units
Sold at
International Distribution only in countries of: Germany and Switzerland.
Description
The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery.
Product
TC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 6 RIGHT, STERILE R, REF 12 000 0011, S&N 7500266, Smith & Nephew Orthopaedics AG Product Usage: Tibial knee prosthesis
Manufacturer(s)
Smith & Nephew, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →