RecallWatchUSA

Home / Recalls / Medical Devices

Smiths Medical ASD, Inc. recalls CADD¿- Solis Ambulatory Infusion Pump, Model 2111 & 2112, Version 1.0, Version 2.0 and Version 3.0 & Model 2120, Versio

Recalled June 24, 2015 · FDA recall Z-1813-2015 · Smiths Medical ASD

Hazard
Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD¿-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD¿-Solis Medication Cassette Reservoir or Administration Set disposable to the pump. When binding occurs, it can prevent the key from fully rotating the Cassette/ Keypad Lock to the locked (e · FDA Class II
Units
4426 (2427 US, 1999 OUS)
Sold at
Worldwide Distribution-US (nationwide) including the states of AL, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, LA, MA, MI, MN, MS, MO, NE, NM, NY, NC, ND, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, and WI, and the countries of AT (Austria), AU (Australia), BE (Belgium), CA (Canada), CH (Switzerland), DE (Germany), DK (Denmark), FR (France), GB (Great Britain), HK (Hong Kong), LB (Lebanon), NL (Netherlands), NZ (New Zealand), QA (Qatar), SA (Saudi Arabia), and SE (Sweden).

Description

Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD¿-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD¿-Solis Medication Cassette Reservoir or Administration Set disposable to the pump. When binding occurs, it can prevent the key from fully rotating the Cassette/ Keypad Lock to the locked (e

Product
CADD¿- Solis Ambulatory Infusion Pump, Model 2111 & 2112, Version 1.0, Version 2.0 and Version 3.0 & Model 2120, Version 1.0. Provide measured drug therapy to patients in hospital or outpatient settings.
Manufacturer(s)
Smiths Medical ASD, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Smiths Medical ASD, Inc. recalls CADD¿- Solis Ambulatory Infusion Pump, Model 2111 & 2112, Version 1.0, Version 2.0 and Version 3.0 & Model 2120, Versio | RecallWatchUSA