RecallWatchUSA

Home / Recalls / Medical Devices

Smiths Medical ASD Inc. recalls DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port f

Recalled May 24, 2023 · FDA recall Z-1580-2023 · Smiths Medical ASD

Hazard
DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a single patient. · FDA Class II
Units
47007 units
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, New Zealand, and Hong Kong.

Description

DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a single patient.

Product
DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port for use with breathing circuits
Manufacturer(s)
Smiths Medical ASD Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov