Smiths Medical ASD Inc. recalls Jelco, Hypodermic needle Pro¿ needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only
Recalled July 12, 2017 · FDA recall Z-2670-2017 · Smiths Medical ASD
Hazard
Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro¿ needles, Smiths Medical part numbers EL 1915 and SB5021. · FDA Class II
Units
736,000
Sold at
US Distribution to the state of : IN
Description
Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro¿ needles, Smiths Medical part numbers EL 1915 and SB5021.
Product
Jelco, Hypodermic needle Pro¿ needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO.
Manufacturer(s)
Smiths Medical ASD Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →