Home / Recalls / Medical Devices
Smiths Medical Asd Inc recalls Medex LOGICAL CATH LAB KIT, List Number M20754
Recalled January 24, 2024 · FDA recall Z-0739-2024 · Smiths Medical Asd
Hazard
There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication. · FDA Class II
Units
31,685 total
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, UAE.
Description
There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
Product
Medex LOGICAL CATH LAB KIT, List Number M20754
Manufacturer(s)
Smiths Medical Asd Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →