RecallWatchUSA

Home / Recalls / Medical Devices

Smiths Medical ASD Inc. recalls Medfusion Syringe Pump, Model 4000. The Model 4000 pump is a small, lightweight and portable syringe infusion pump. Th

Recalled December 18, 2019 · FDA recall Z-0610-2020 · Smiths Medical ASD

Hazard
There is an anomaly in Medfusion 4000 Syringe Pump Firmware version 1.7.0 that could potentially result in loss of primary audible and visual battery alarm functionality, and interruption of therapy. · FDA Class I
Units
626 devices
Sold at
CO, MD, NY

Description

There is an anomaly in Medfusion 4000 Syringe Pump Firmware version 1.7.0 that could potentially result in loss of primary audible and visual battery alarm functionality, and interruption of therapy.

Product
Medfusion Syringe Pump, Model 4000. The Model 4000 pump is a small, lightweight and portable syringe infusion pump. The pump offers a variety of delivery modes programmable to meet specific patient care needs.
Manufacturer(s)
Smiths Medical ASD Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov