Home / Recalls / Medical Devices
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 2.5mm TTS CUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/L
Recalled September 4, 2024 · FDA recall Z-2615-2024 · Smiths Medical ASD
Hazard
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect. · FDA Class I
Sold at
Worldwide distribution.
Description
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Product
PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 2.5mm TTS CUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SN025; b) 3.0mm TTS CUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SN030; c) 3.5mm TTS CUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SN035; d) 4.0mm TTS CUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SN040; e) 2.5mm TTS CUFFED PEDIATRIC TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SP025; f) 3.0mm TTS CUFFED PEDIATRIC TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SP030; g) 3.5mm TTS CUFFED PEDIATRIC TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SP035; h) 4.0mm TTS CUFFED PEDIATRIC TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SP040; i) 4.5mm TTS CUFFED PEDIATRIC TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SP045; j) 5.0mm TTS CUFFED PEDIATRIC TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SP050; k) 5.5mm TTS CUFFED PEDIATRIC TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SP055
Manufacturer(s)
Smiths Medical ASD Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →