Home / Recalls / Medical Devices
Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-5
Recalled April 24, 2024 · FDA recall Z-1541-2024 · Smiths Medical ASD
Hazard
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded. · FDA Class II
Units
7160 pumps
Sold at
Worldwide distribution.
Description
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Product
Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e) 4000-0105-50, f) 4000-0105-51, g) 4000-0105-78, h) 4000-0106-00, i) 4000-0106-01, j) 4000-0106-231,
Manufacturer(s)
Smiths Medical ASD Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →