RecallWatchUSA

Home / Recalls / Medical Devices

Solta Medical Inc recalls VASER PRO Amplifier, VASER 2.2 VASERlipo System stem, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intend

Recalled August 30, 2017 · FDA recall Z-3048-2017 · Solta Medical

Hazard
Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017. · FDA Class II
Units
7 units
Sold at
US Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.

Description

Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.

Product
VASER PRO Amplifier, VASER 2.2 VASERlipo System stem, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).
Manufacturer(s)
Solta Medical Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov