RecallWatchUSA

Home / Recalls / Medical Devices

Sophysa S.A. recalls The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial

Recalled May 22, 2024 · FDA recall Z-1825-2024 · Sophysa S.A

Hazard
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion. · FDA Class II
Units
3487 units OUS
Sold at
Worldwide - US Nationwide distribution in the states of AZ, NY, TX and the countries of ARGENTINA, CHINA, CZECH REPUBLIC, FRANCE, GREECE, HUNGARY, JAPAN, PAKISTAN, Portugal, ROMANIA, SOUTH AFRICA, SWEDEN, VIETNAM.

Description

An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.

Product
The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VT
Manufacturer(s)
Sophysa S.A.
Made in
France