Home / Recalls / Medical Devices
Spacelabs Healthcare, Inc. recalls Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Recalled April 15, 2026 · FDA recall Z-1793-2026 · Spacelabs Healthcare
Hazard
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload. · FDA Class II
Units
1790
Sold at
Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
Description
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Product
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Manufacturer(s)
Spacelabs Healthcare, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →