RecallWatchUSA

Home / Recalls / Medical Devices

Specialty Appliance Works, Inc. recalls LR Telescoping Cantilever Arms (5/pk) and LL Telescoping Cantilever Arms (5/pk) A Herbst Appliance is used to affect ch

Recalled April 29, 2015 · FDA recall Z-1508-2015 · Specialty Appliance Works

Hazard
The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS. · FDA Class II
Units
14,428 Universal Nuts
Sold at
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.

Description

The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.

Product
LR Telescoping Cantilever Arms (5/pk) and LL Telescoping Cantilever Arms (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
Manufacturer(s)
Specialty Appliance Works, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Specialty Appliance Works, Inc. recalls LR Telescoping Cantilever Arms (5/pk) and LL Telescoping Cantilever Arms (5/pk) A Herbst Appliance is used to affect ch | RecallWatchUSA