RecallWatchUSA

Home / Recalls / Medical Devices

SpineFrontier, Inc. recalls Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to pro

Recalled August 14, 2013 · FDA recall Z-1871-2013 · SpineFrontier

Hazard
Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference. · FDA Class II
Units
13 drill guides
Sold at
US Distribution including Puerto Rico and the states of CO, KS, MA and TX.

Description

Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference.

Product
Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.
Manufacturer(s)
SpineFrontier, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

SpineFrontier, Inc. recalls Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to pro | RecallWatchUSA