RecallWatchUSA

Home / Recalls / Medical Devices

SpineFrontier, Inc. recalls Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of i

Recalled August 21, 2013 · FDA recall Z-1957-2013 · SpineFrontier

Hazard
The nipple of the tipped driver broke off when surgeons were attempting to insert the screw. · FDA Class II
Units
9
Sold at
Distributed in DC and the states of MA, and OK.

Description

The nipple of the tipped driver broke off when surgeons were attempting to insert the screw.

Product
Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of insertion during fixation of spinal intervertebral.
Manufacturer(s)
SpineFrontier, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

SpineFrontier, Inc. recalls Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of i | RecallWatchUSA