Home / Recalls / Medical Devices
St. Jude Medical, Atrial Fibrillation Division, Inc. recalls TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.
Recalled January 31, 2024 · FDA recall Z-0814-2024 · St. Jude Medical, Atrial Fibrillation Division
Hazard
When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted. · FDA Class II
Units
483 units
Sold at
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
Description
When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.
Product
TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.
Manufacturer(s)
St. Jude Medical, Atrial Fibrillation Division, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →