Recalled March 4, 2020 · FDA recall Z-1351-2020 · St Jude Medical, Cardiac Rhythm Management Division
Hazard
Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with wireless RF telemetry or monitored remotely. · FDA Class II
Units
256
Sold at
U.S.: WV, MS, TX, DE, OH, AZ, MI, NC, KY, WA, NY, CA, IN, MO, MD, NJ, FL, PA, OK. O.U.S.: United Kingdom, Turkey, Sweden, Spain, Portugal, Poland, Philippines, Pakistan, Norway, New Zealand, Netherlands, Japan, Italy, Hong Kong, Greece, Germany, France, Finland, Denmark, Canada, Belgium, Austria, Australia
Description
Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with wireless RF telemetry or monitored remotely.
Product
Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (05414734507585), CD1277-36 (05414734506045), CD1377-36C (05414734507622), CD2277-36Q (05414734505932), CD2377-36QC (05414734507523), with Merlin Patient Care System with software model 3330