RecallWatchUSA

Home / Recalls / Medical Devices

St Jude Medical, Cardiac Rhythm Management Division recalls Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (0541473450

Recalled March 4, 2020 · FDA recall Z-1351-2020 · St Jude Medical, Cardiac Rhythm Management Division

Hazard
Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with wireless RF telemetry or monitored remotely. · FDA Class II
Units
256
Sold at
U.S.: WV, MS, TX, DE, OH, AZ, MI, NC, KY, WA, NY, CA, IN, MO, MD, NJ, FL, PA, OK. O.U.S.: United Kingdom, Turkey, Sweden, Spain, Portugal, Poland, Philippines, Pakistan, Norway, New Zealand, Netherlands, Japan, Italy, Hong Kong, Greece, Germany, France, Finland, Denmark, Canada, Belgium, Austria, Australia

Description

Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with wireless RF telemetry or monitored remotely.

Product
Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (05414734507585), CD1277-36 (05414734506045), CD1377-36C (05414734507622), CD2277-36Q (05414734505932), CD2377-36QC (05414734507523), with Merlin Patient Care System with software model 3330
Manufacturer(s)
St Jude Medical, Cardiac Rhythm Management Division

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

St Jude Medical, Cardiac Rhythm Management Division recalls Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (0541473450 | RecallWatchUSA