St. Jude Medical, Cardiac Rhythm Management Division recalls Merlin PCS 3650 programmer Model 3330 software

Recalled April 27, 2022 · FDA recall Z-0964-2022 · St. Jude Medical, Cardiac Rhythm Management Division

Hazard
Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters. · FDA Class II
Units
30,069 devices with software
Sold at
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN. IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY. The countries of Argentina, Australia, AUSTRIA, Bahrain, Bangladesh, BELGIUM, Brazil, Bulgaria, Canada, Chile, COLOMBIA, COSTA RICA, Cyprus, Czech republic, DENMARK, ECUADOR, Egypt, Estonia, Faroe Islands, FINLAND, FRANCE, French Guiana, French Polynesia, GERMANY, Greece, Greenland, Guadeloupe, HUNGARY, Iceland, India, IRELAND, ITALY, Japan, Jordan, Korea, Kuwait, Latvia, Lebanon, Liechtenstein, LITHUANIA, Luxembourg, Malaysia, Malta, Martinque, Mayotte, MEXICO, Nepal, NETHERLANDS, New Caledonia, New Zealand, NORWAY, Oman, Pakistan, Palestine, POLAND, PORTUGAL, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, SPAIN, Srilanka, SWEDEN, SWITZERLAND, Thailand, UAE, and UNITED KINGDOM.

Description

Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.

Product
Merlin PCS 3650 programmer Model 3330 software
Manufacturer(s)
St. Jude Medical, Cardiac Rhythm Management Division