RecallWatchUSA

Home / Recalls / Medical Devices

St. Jude Medical, Cardian Rhythm Management Division recalls ENDURITY Pulse Generator REF PM**** SN ********* St. Jude Medical Cardiac Rhythm Assurity and Endurity pacemakers

Recalled May 19, 2021 · FDA recall Z-1531-2021 · St. Jude Medical, Cardian Rhythm Management Division

Hazard
Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough · FDA Class I
Units
95,042 pacemakers [updated 03/20/2023]
Sold at
Worldwide distribution: U.S (nationwide) to states including.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.( countries): [806 Amendment - after March 2021 - updated 3/20/2023] Algeria, Angola, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Benin, Bolivia, Brazil, Bulgaria, Cameroon, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Guatemala, Hong Kong, Hungary, India, Ireland, Italy, Japan, Jordan, Korea, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Mexico, Montenegro, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Palestine, Panama, Paraguay, Peru Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Senegal, Servia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Yemen.

Description

Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough

Product
ENDURITY Pulse Generator REF PM**** SN ********* St. Jude Medical Cardiac Rhythm Assurity and Endurity pacemakers are intended to provide single-chamber or dual-chamber bradycardia pacing.
Manufacturer(s)
St. Jude Medical, Cardian Rhythm Management Division

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

St. Jude Medical, Cardian Rhythm Management Division recalls ENDURITY Pulse Generator REF PM**** SN ********* St. Jude Medical Cardiac Rhythm Assurity and Endurity pacemakers | RecallWatchUSA