Home / Recalls / Medical Devices
St Jude Medical Inc. recalls Fortify VR, Model No. CD1231-40, CD1231-40Q. Fortify ST VR, Model No. CD1241-40, CD1241-40Q. Fortify Assura VR, Model
Recalled November 2, 2016 · FDA recall Z-0115-2017 · St Jude Medical
Hazard
St Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit. · FDA Class I
Units
204,785 units
Sold at
Worldwide Distribution - US (nationwide) and OUS
Description
St Jude Medical is recalling implantable cardiac defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) devices because they are subject to premature battery depletion due to lithium cluster with the battery causing a short circuit.
Product
Fortify VR, Model No. CD1231-40, CD1231-40Q. Fortify ST VR, Model No. CD1241-40, CD1241-40Q. Fortify Assura VR, Model No. CD1257-40, CD1257-40Q, CD1357-40C. Fortify Assura ST VR, Model No. CD1263-40, CD1263-40Q, CD1363-40, CD1363-40Q. Fortify DR, Model No. CD2231-40, CD2231-40Q. Fortify ST DR, Model No. CD2241-40, CD-2241-40Q, CD2263-40, CD2363-40C, CD2363-40Q. Fortify Assura DR, Model No. CD2257-40, CD2257-40Q, CD2357-40C, CD2357-40Q. Fortify and Fortify Assura implantable cardioverter defibrillators (ICDS) are intended to provided ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing. Unify and Unify Assura cardiac resynchronization therapy defibrillators (CRTDs) are intended to provide ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing, in additional to cardiac resynchronization therapy via left ventricular pacing
Manufacturer(s)
St Jude Medical Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →