Recalled October 25, 2017 · FDA recall Z-0023-2018 · St. Jude Medical

Hazard
Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations. · FDA Class II
Units
3,119 devices
Sold at
Worldwide Distribution - US (nationwide) (foreign consignee list will be provide by the firm at a later time)

Description

Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.

Product
Infinity 5 Implantable Pulse Generator and Infinity 7 Implantable Pulse Generator Models 6660, 6661, 6662, 6663
Manufacturer(s)
St. Jude Medical, Inc.