St Jude Medical Inc. recalls Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to su

Recalled July 11, 2018 · FDA recall Z-2335-2018 · St Jude Medical

Hazard
The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal. · FDA Class II
Units
0
Sold at
U.S. Nationwide Distribution: PR, MI, NJ, IL, FL, AR, PA, KY, CA, TN, IN, AL, TX, AZ, SC, OK, VA, NH, NM, WA, CO, IA, WI, NV, UT, OH, NC, MO, DE, WV, MD, OR, NE, VT, KS, MA, IA, WY, CT, GA, MS, LA, DC, LA,

Description

The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.

Product
Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.
Manufacturer(s)
St Jude Medical Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

St Jude Medical Inc. recalls Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to su | RecallWatchUSA