Stanbio Laboratory, LP recalls STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.
Recalled December 13, 2023 · FDA recall Z-0490-2024 · Stanbio Laboratory
Hazard
Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis. · FDA Class II
Units
2,351
Sold at
Worldwide - US Nationwide distribution including in the states of AZ, RI, MO, UT, AK, LA, TX, OH, GA, MN, NY, CA, NJ, IL, WA, KS, MI, NC, PA, MA, CO, KY, NM, FL, MD, VA, IN, AL, SD, MS, WV, WI, ID, MT, WY and the countries of Canada and Singapore.
Description
Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.
Product
STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.
Manufacturer(s)
Stanbio Laboratory, LP
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →