Home / Recalls / Medical Devices
Stanmore Implants Worldwide Ltd. recalls Tibial Component Metal Cased Small Long; Tibial Component Metal Cased Small Short; Tibial Component Metal Cased Std Long
Recalled December 28, 2016 · FDA recall Z-0821-2017 · Stanmore Implants Worldwide
Hazard
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem. · FDA Class II
Units
51 units
Sold at
Nationwide Distribution to OH, NE, FL, LA, NY, TX, MS, MD, CT, AL, PA
Description
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
Product
Tibial Component Metal Cased Small Long; Tibial Component Metal Cased Small Short; Tibial Component Metal Cased Std Long; Tibial Component Metal Cased Std Short The product is a component is a systems that is intended for the replacement if diseased or deficient bone around the knee joint, distal femur, proximal femur or total femur bone
Manufacturer(s)
Stanmore Implants Worldwide Ltd.
Made in
United Kingdom