RecallWatchUSA

Home / Recalls / Medical Devices

Stelkast Co recalls EXp Tibial Insert ; Intended Use of Device(s): The Proven Gen-Flex Total Knee System is intended for: a. Total knee rep

Recalled January 21, 2015 · FDA recall Z-0970-2015 · Stelkast

Hazard
Difficulty locking EXp tibial inserts into tibial trays, caused by a manufacturing condition where a lead in radius on the anterior locking tab is not present on certain insert lots. The absence of this lead in may prevent or make assembly of the insert and tray difficult. · FDA Class II
Units
553
Sold at
TX, MD, IN, KS, CA, PA, GA, WA, NY, IL.

Description

Difficulty locking EXp tibial inserts into tibial trays, caused by a manufacturing condition where a lead in radius on the anterior locking tab is not present on certain insert lots. The absence of this lead in may prevent or make assembly of the insert and tray difficult.

Product
EXp Tibial Insert ; Intended Use of Device(s): The Proven Gen-Flex Total Knee System is intended for: a. Total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis, and/or post-traumatic degenerative problems; and b. Revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.
Manufacturer(s)
Stelkast Co

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stelkast Co recalls EXp Tibial Insert ; Intended Use of Device(s): The Proven Gen-Flex Total Knee System is intended for: a. Total knee rep | RecallWatchUSA