Steris Corporation recalls AMSCO 5000 Series Washer/Disinfector Model # 5052
Recalled March 20, 2019 · FDA recall Z-0996-2019 · Steris
Hazard
The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection. · FDA Class II
Units
68
Sold at
The products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, and Taiwan.
Description
The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
Product
AMSCO 5000 Series Washer/Disinfector Model # 5052
Manufacturer(s)
Steris Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →