Home / Recalls / Medical Devices
Steris Corporation recalls Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062
Recalled May 24, 2023 · FDA recall Z-1595-2023 · Steris
Hazard
In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire. · FDA Class II
Units
322 units
Sold at
Worldwide Distribution: US (Nationwide) and OUS (Foreign) to countries of: China, Lithuania, Saudi Arabia, Korea, Republic of, Canada, Taiwan, United Kingdom, Spain, Australia, New Zealand, Spain, Brazil, Bahrain, Switzerland, Great Britain, South Africa, Hong Kong, Qatar.
Description
In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Product
Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062
Manufacturer(s)
Steris Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →