Home / Recalls / Medical Devices
Steris Corporation recalls Reliance Vision Single-Chamber Washer/Disinfector - Product Usage: intended for use in the cleaning and intermediate lev
Recalled January 27, 2021 · FDA recall Z-0876-2021 · Steris
Hazard
The electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/Disinfector may malfunction, potentially leading to the heating elements in the drying chamber to overheat, eventually resulting in smoke and/or fire. · FDA Class II
Units
3405
Sold at
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Bulgaria, Canada, Chile, China, Ecuador, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of Malaysia, Mexico, New Zealand, Panama, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela.
Description
The electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/Disinfector may malfunction, potentially leading to the heating elements in the drying chamber to overheat, eventually resulting in smoke and/or fire.
Product
Reliance Vision Single-Chamber Washer/Disinfector - Product Usage: intended for use in the cleaning and intermediate level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments, such as forceps and clamps, theatre shoes, and other similar and related articles found in healthcare facilities.
Manufacturer(s)
Steris Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →