Home / Recalls / Medical Devices
Steris Corporation recalls Steris Lighting and Visualization Systems, Sterile Light handle Cover LB44 -Sterilizable consumables (disposable, sing
Recalled May 10, 2023 · FDA recall Z-1476-2023 · Steris
Hazard
Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient · FDA Class II
Units
263,280 covers
Sold at
Worldwide - US Nationwide distribution including Puerto Rico and the countries of Canada, Australia.
Description
Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient
Product
Steris Lighting and Visualization Systems, Sterile Light handle Cover LB44 -Sterilizable consumables (disposable, single use) used to drape the lighthandle of STERIS surgical lighting systems for the purpose of maintaining a sterile surgical environment. REF: LB44
Manufacturer(s)
Steris Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →