RecallWatchUSA

Home / Recalls / Medical Devices

Steris Corporation recalls SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Li

Recalled January 8, 2014 · FDA recall Z-0616-2014 · Steris

Hazard
Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts. · FDA Class II
Units
6,028 units
Sold at
Nationwide Distribution - USA (nationwide) and the District of Columbia and Puerto Rico. Product was also shipped to the following countries: Afghanistan, Bahrain, Hong Kong, Italy, Saudi Arabia, Turkey & United Kingdom.

Description

Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.

Product
SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.
Manufacturer(s)
Steris Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Steris Corporation recalls SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Li | RecallWatchUSA