RecallWatchUSA

Home / Recalls / Medical Devices

Steritec Products Mfc. Co., Inc. recalls Getinge Assure Accufast BI Test Pack w/ 5 Controls, SteriTec Product Code PL 395, Distributor Product Code 61301606638

Recalled August 7, 2013 · FDA recall Z-1825-2013 · Steritec Products Mfc

Hazard
Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims. · FDA Class II
Units
10,170 units
Sold at
Worldwide distribution: US (Nationwide) and countries of: Argentina, Australia, Canada, British Virgin Island, Czech Republic, Italy, and Turkey.

Description

Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.

Product
Getinge Assure Accufast BI Test Pack w/ 5 Controls, SteriTec Product Code PL 395, Distributor Product Code 61301606638 The Smart-Read Biological Test Pack with Instant Readout Integrator: is designed specifically for biological testing of 3 or 4 minutes or longer at 132 (270 and 3 minutes at 135 pre-vacuum steam sterilizers.
Manufacturer(s)
Steritec Products Mfc. Co., Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov