RecallWatchUSA

Home / Recalls / Medical Devices

Straumann Manufacturing, Inc. recalls Straumann Twist Drill PRO ¿4.2mm,long,41mm,SS-Non sterile. Dental implant accessory Article Number: 044.255 Dental

Recalled January 18, 2017 · FDA recall Z-0958-2017 · Straumann Manufacturing

Hazard
Drills cannot be inserted into handpiece due to tolerance specification not met · FDA Class II
Units
120 units
Sold at
Nationwide Distribution

Description

Drills cannot be inserted into handpiece due to tolerance specification not met

Product
Straumann Twist Drill PRO ¿4.2mm,long,41mm,SS-Non sterile. Dental implant accessory Article Number: 044.255 Dental implants drill accessory is intended to prepare osteotomy site for surgical dental implant placement. Dental implants are intended for immediate placement and function on single-tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function
Manufacturer(s)
Straumann Manufacturing, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Straumann Manufacturing, Inc. recalls Straumann Twist Drill PRO ¿4.2mm,long,41mm,SS-Non sterile. Dental implant accessory Article Number: 044.255 Dental | RecallWatchUSA