Home / Recalls / Medical Devices
Stryker Communications recalls Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Recalled April 8, 2020 · FDA recall Z-1602-2020 · Stryker Communications
Hazard
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients. · FDA Class II
Units
1415 Surgical Lights
Sold at
US: Nationwide OUS: Canada
Description
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.
Product
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Manufacturer(s)
Stryker Communications
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →