Stryker Communications recalls Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of

Recalled June 14, 2017 · FDA recall Z-2248-2017 · Stryker Communications

Hazard
The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury. · FDA Class II
Units
57 units
Sold at
Worldwide Distribution - US and Canada

Description

The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.

Product
Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated
Manufacturer(s)
Stryker Communications

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov