RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Communications recalls S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected with

Recalled March 31, 2021 · FDA recall Z-1297-2021 · Stryker Communications

Hazard
The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use. · FDA Class II
Units
1,393 units
Sold at
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY. The country of Canada.

Description

The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.

Product
S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD
Manufacturer(s)
Stryker Communications

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Communications recalls S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected with | RecallWatchUSA