Home / Recalls / Medical Devices
Stryker Corporation recalls Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and Nasopore 4cm Firm 2PK Catalog Nu
Recalled April 9, 2025 · FDA recall Z-1480-2025 · Stryker
Hazard
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached. · FDA Class II
Units
182,344 total units
Sold at
US Nationwide distribution.
Description
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
Product
Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and Nasopore 4cm Firm 2PK Catalog Number 5400-222-004S
Manufacturer(s)
Stryker Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →