RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Corporation recalls Triton Canister Software Catalog Number: GAUS-2. Intended adjunct in the estimation of blood loss.

Recalled March 8, 2023 · FDA recall Z-1186-2023 · Stryker

Hazard
Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH. · FDA Class II
Units
88 units
Sold at
US Nationwide distribution.

Description

Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.

Product
Triton Canister Software Catalog Number: GAUS-2. Intended adjunct in the estimation of blood loss.
Manufacturer(s)
Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Corporation recalls Triton Canister Software Catalog Number: GAUS-2. Intended adjunct in the estimation of blood loss. | RecallWatchUSA