RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Endoscopy recalls Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shav

Recalled July 2, 2014 · FDA recall Z-1872-2014 · Stryker Endoscopy

Hazard
STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing. · FDA Class II
Units
1992 trays
Sold at
Worldwide Distribution.

Description

STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.

Product
Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.
Manufacturer(s)
Stryker Endoscopy

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Endoscopy recalls Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shav | RecallWatchUSA