Home / Recalls / Medical Devices
Stryker GmbH recalls Femoral Nail, RIGHT T2 GTN ¿9x320 mm, Product Number 18510932S
Recalled July 18, 2018 · FDA recall Z-2405-2018 · Stryker
Hazard
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage. · FDA Class II
Sold at
The products were distributed to the following US states: AL, Co, FL, KY, NC, TN, UT, and WA.
Description
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
Product
Femoral Nail, RIGHT T2 GTN ¿9x320 mm, Product Number 18510932S
Manufacturer(s)
Stryker GmbH
Made in
Switzerland
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →