Home / Recalls / Medical Devices
Stryker GmbH recalls Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002
Recalled October 3, 2018 · FDA recall Z-3240-2018 · Stryker
Hazard
When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan. · FDA Class II
Units
233
Sold at
Nationwide
Description
When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.
Product
Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002
Manufacturer(s)
Stryker GmbH
Made in
Switzerland
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →