Home / Recalls / Medical Devices
Stryker Howmedica Osteonics Corp. recalls Stryker Howmedica Osteonics; Stryker T2 Humerus System Teflon Tube, sterile; REF 1806-0073S The T2 Nailing System is
Recalled January 2, 2013 · FDA recall Z-0610-2013 · Stryker Howmedica Osteonics
Hazard
During the manufacturing of the affected lot a manufacturing error has led to embrittlement of the tube material of the T2 Humerus Teflon Tube. · FDA Class II
Units
100 units (73 units US; 27 units OUS)
Sold at
Worldwide distribution: USA (nationwide) including states of: AL, CO, CT, IL, IN, MO, MS, NC, NJ, NY, OR,and TX; and countries of: Australia, Canada, France, Spain and United Kingdom.
Description
During the manufacturing of the affected lot a manufacturing error has led to embrittlement of the tube material of the T2 Humerus Teflon Tube.
Product
Stryker Howmedica Osteonics; Stryker T2 Humerus System Teflon Tube, sterile; REF 1806-0073S The T2 Nailing System is designed to treat complex fractures in Humeral, Femoral, Tibial and Ankle applications. The Teflon Tube is an optional instrument used to exchange the Ball Tip Guide Wire with the Smooth Tip Guide Wire once reaming is completed. Stryker Stryker Trauma GmbH 24232 Schonkirchen, Germany distributed in USA by: Howmedica Osteonics Corp., 235 Corporate Drive,
Manufacturer(s)
Stryker Howmedica Osteonics Corp.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →