Stryker Howmedica Osteonics Corp. recalls Stryker Orthopaedics TIBIAL OFFSET BUSHING TRIATHLON REVISION INSTRUMENTS REF 6543-2-600;NON-STERILE Howmedica O
Recalled December 12, 2012 · FDA recall Z-0470-2013 · Stryker Howmedica Osteonics
Hazard
Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings. · FDA Class II
Units
8
Sold at
Worldwide Distribution.
Description
Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings.
Product
Stryker Orthopaedics TIBIAL OFFSET BUSHING TRIATHLON REVISION INSTRUMENTS REF 6543-2-600;NON-STERILE Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA A Subsidiary of Stryker Corp. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 Meyzieu Cedex France Made in USA The Offset Bushings are parts of the Triathlon TS system. They are used during trialing to assess the appropriate magnitude and orientation of offset between the Implant and the intramedullary Stem. The feedback from the Offset Bushing allows for the selection of the corresponding Offset Adaptor, which allows for optimal bone coverage, compensating for differences between the Implant and bone canal axes.
Manufacturer(s)
Stryker Howmedica Osteonics Corp.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →