RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Instruments Div. of Stryker Corporation recalls RF Cannula, 100 mm,Curved 5 MM, 20 G, REF 0406-630-115 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF

Recalled August 8, 2012 · FDA recall Z-2134-2012 · Stryker Instruments Div. of Stryker

Hazard
Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead of 5mm and the cannula is used in a cervical procedure there is a potential for a thermal skin lesion due to a larger lesion creation. If the 10 mm active tip cannula, as opposed · FDA Class II
Units
210
Sold at
Worldwide Distribution - US Nationwide and the countries of England, France, Netherlands, and Switzerland.

Description

Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead of 5mm and the cannula is used in a cervical procedure there is a potential for a thermal skin lesion due to a larger lesion creation. If the 10 mm active tip cannula, as opposed

Product
RF Cannula, 100 mm,Curved 5 MM, 20 G, REF 0406-630-115 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.
Manufacturer(s)
Stryker Instruments Div. of Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Instruments Div. of Stryker Corporation recalls RF Cannula, 100 mm,Curved 5 MM, 20 G, REF 0406-630-115 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF | RecallWatchUSA