Stryker Instruments Div. of Stryker Corporation recalls Stryker Instruments 0400-850-100 T5 zipper Toga with Peel Away, Sterile, EO; single use; Product Usage: Protection

Recalled June 12, 2013 · FDA recall Z-1504-2013 · Stryker Instruments Div. of Stryker

Hazard
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr · FDA Class II
Units
6,990 units ( 699 boxes)
Sold at
USA Nationwide Distribution

Description

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

Product
Stryker Instruments 0400-850-100 T5 zipper Toga with Peel Away, Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.
Manufacturer(s)
Stryker Instruments Div. of Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov