RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Instruments Div. of Stryker Corporation recalls Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is us

Recalled April 30, 2014 · FDA recall Z-1493-2014 · Stryker Instruments Div. of Stryker

Hazard
During packaging verification testing, a failure occurred due to cuts in the packaging potentially causing a breach in the sterile barrier. Additional testing identified microscopic aluminum metal shavings and foreign particles estimated in size to be between 0.01inches x 0.015 inches to 0.03 inches x 0.04 inches that were introduced during the manufacturing process and were sterilized within the · FDA Class II
Units
35,310 unexpired units
Sold at
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV,and WY, and the countries of Austria, Canada, Sweden, England, UAE, Netherlands, Spain, and Hong Kong.

Description

During packaging verification testing, a failure occurred due to cuts in the packaging potentially causing a breach in the sterile barrier. Additional testing identified microscopic aluminum metal shavings and foreign particles estimated in size to be between 0.01inches x 0.015 inches to 0.03 inches x 0.04 inches that were introduced during the manufacturing process and were sterilized within the

Product
Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is used to remove smoke, particles, and body and casual fluids from the point of surgical activity during surgical procedures that use ESU for cutting and cauterizing.
Manufacturer(s)
Stryker Instruments Div. of Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Instruments Div. of Stryker Corporation recalls Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is us | RecallWatchUSA