RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Instruments Div. of Stryker Corporation recalls SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting, displaying, re

Recalled July 27, 2022 · FDA recall Z-1452-2022 · Stryker Instruments Div. of Stryker

Hazard
When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a case note and allowing the case to close. The software update will prevent the occurrence of this error. · FDA Class II
Units
44 units
Sold at
The products were distributed to the following US states: KS and ME.

Description

When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a case note and allowing the case to close. The software update will prevent the occurrence of this error.

Product
SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting, displaying, recording and locating the number of RFID-tagged absorbent articles used during surgical procedures.
Manufacturer(s)
Stryker Instruments Div. of Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Instruments Div. of Stryker Corporation recalls SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting, displaying, re | RecallWatchUSA