Home / Recalls / Medical Devices
Stryker Instruments Div. of Stryker Corporation recalls The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite Camera, SPC-3 Com
Recalled June 11, 2014 · FDA recall Z-1716-2014 · Stryker Instruments Div. of Stryker
Hazard
The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints · FDA Class II
Units
20
Sold at
Distributed in the states of AR, CA, IN, LA, MA, MD, MI, MT, NV, and TX.
Description
The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints
Product
The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite Camera, SPC-3 Computer, IO-Tablet (touch screen) and the monitor).
Manufacturer(s)
Stryker Instruments Div. of Stryker Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →