RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Instruments recalls Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.

Recalled August 3, 2022 · FDA recall Z-1478-2022 · Stryker Instruments

Hazard
There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures. · FDA Class II
Units
543
Sold at
Worldwide distribution - US Nationwide distribution in the states of Arizona, California, Florida, Georgia, Iowa, Michigan, Missouri, New Mexico, North Carolina, Ohio, Texas, Virginia and the countries of Canada, Netherlands, and Thailand.

Description

There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.

Product
Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.
Manufacturer(s)
Stryker Instruments

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov