Home / Recalls / Medical Devices
Stryker Leibinger GmbH & Co. KG recalls 8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution
Recalled May 29, 2024 · FDA recall Z-1885-2024 · Stryker Leibinger GmbH & Co. KG
Hazard
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction. · FDA Class II
Units
705 units
Sold at
US Nationwide distribution.
Description
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
Product
8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.
Manufacturer(s)
Stryker Leibinger GmbH & Co. KG
Made in
Germany
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →