Home / Recalls / Medical Devices
Stryker Leibinger GmbH & Co. KG recalls MEDPOR Titan OFW - MTB - Right - 1.0mm, Catalog number 81032
Recalled March 9, 2022 · FDA recall Z-0715-2022 · Stryker Leibinger GmbH & Co. KG
Hazard
Inside the packaging of one reported item a non-conforming product was found. Contrary to the design, the barrier layer was manufactured on the inferior side of the implant versus the superior side. · FDA Class II
Units
74
Sold at
Worldwide distribution - US Nationwide.
Description
Inside the packaging of one reported item a non-conforming product was found. Contrary to the design, the barrier layer was manufactured on the inferior side of the implant versus the superior side.
Product
MEDPOR Titan OFW - MTB - Right - 1.0mm, Catalog number 81032
Manufacturer(s)
Stryker Leibinger GmbH & Co. KG
Made in
Germany
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →